Skip to main content

Table 2 Baseline Lesion Characteristics

From: Diagnostic performance of target vs. vessel μFR in stable coronary artery disease

Lesion Characteristics Number of vessels (146)

target-μFR

vessel-μFR

P

QCA feature

 Diameter stenosis, %

67.64 ± 10.174

-

 LAD

92 (63.0%)

-

 LCX

18 (12.3%)

-

 RCA

36 (24.7%)

-

QFR feature

 Reference vessel diameter, mm

2.94(2.44, 3.29)

2.80(2.34, 3.19)

P > 0.05

 Minimal lumen diameter, mm

1.61 ± 0.51

1.55 ± 0.47

P > 0.05

 Diameter stenosis (%)

45.1 ± 10.5

44.2 ± 10.7

P > 0.05

 Area Stenosis (%)

2.91 ± 0.65

2.78 ± 0.61

P > 0.05

Functional indexes

 target-μFR value

0.83 ± 0.101

-

 vessel-μFR value

0.83 ± 0.098

-

 FFR

0.82 ± 0.098

-

 FFR ≤ 0.80

61 (41.5%)

-

  1. Values areMean ± SD, Mean (P25,P75) or n (%). LAD lanterior descending coronary artery, LCX left circumflex artery, RCA right coronary artery. QCA quantitative coronary angiography. Target-μFR target-position Murray Law based quantitative flow ratio. vessel-μFR Vessel quantitative flow ratio derived from Murray Law